[INSTRM-1027] b2 gatevalve won't open Created: 25/Jun/20  Updated: 05/Aug/20  Resolved: 16/Jul/20

Status: Won't Fix
Project: Instrument control development
Component/s: None
Affects Version/s: None
Fix Version/s: None

Type: Bug Priority: Blocker
Reporter: arnaud.lefur Assignee: cloomis
Resolution: Won't Fix Votes: 0
Labels: SPS
Remaining Estimate: Not Specified
Time Spent: Not Specified
Original Estimate: Not Specified

Issue Links:
Relates
relates to INSTRM-1028 (re-)Add BEE-side ADIO mask to gateva... Done
Sprint: SM1PD-2020 G

 Description   

we were unable to open B2 gatevalve :

2020-06-24 12:13:03 cmdIn=xcu_b2 gatevalve open underVacuum 
2020-06-24 12:13:03 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 
2020-06-24 12:13:04 xcu_b2 i turbospeed="90000" 
2020-06-24 12:13:04 xcu_b2 i turbostatus=0x000502bc,"Above normal speed, Vent valve energized, Start command active, Serial enable active, Above 50% full speed, Serial control mode, bit 17, bit 19" 
2020-06-24 12:13:04 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 
2020-06-24 12:13:05 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 
2020-06-24 12:13:05 xcu_b2 i gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 
2020-06-24 12:13:09 xcu_b2 w unparsedtext="failed to get desired gate valve state. Timed out with: 0b1110100)" 
2020-06-24 12:13:09 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 
2020-06-24 12:13:09 xcu_b2 :


 Comments   
Comment by cloomis [ 25/Jun/20 ]

The most obviously interesting points are the bit masks from the interlock board and from the ADIO. The gatevalve/ADIO mask shows no change, so it is not clear that a change is being requested. And the interlock board clearly does not see the request at that end.

Comment by cloomis [ 25/Jun/20 ]

With the new interlock board, the gatevalve status is actually the interlock status. So that is not meaningful.

Comment by cloomis [ 01/Jul/20 ]

We swapped BEEs, and the problem was clearly on just one of them: this is a hardware problem. Besides adding the ADIO mask in INSTRM-1028, I do not think there is more to do here – can hassan or fmadec close this if they agree?

Comment by fmadec [ 01/Jul/20 ]

I am ok to close, this is now track in bugzilla :

https://pfs.ipmu.jp/internal/bts/show_bug.cgi?id=807

Comment by cloomis [ 16/Jul/20 ]

Hardware issue, tracked by bugzilla 807