[INSTRM-1018] update rexm switch search routine Created: 19/Jun/20 Updated: 05/Aug/20 Resolved: 01/Jul/20 |
|
| Status: | Done |
| Project: | Instrument control development |
| Component/s: | ics_enuActor |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Task | Priority: | Normal |
| Reporter: | arnaud.lefur | Assignee: | arnaud.lefur |
| Resolution: | Done | Votes: | 0 |
| Labels: | SPS | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Story Points: | 2 |
| Sprint: | SM1PD-2020 G |
| Description |
|
The procedure to accurately position the rda has been updated by winlight.
Now, it will go all the way at 3.2mm, decelerate at 10,24 mm/sec² once the switch is hit, this will give the 500microns to fully engage the switch. |
| Comments |
| Comment by arnaud.lefur [ 01/Jul/20 ] |
|
tested today at LAM |