[INSTRM-1012] red shutter command error Created: 12/Jun/20 Updated: 15/Apr/21 Resolved: 10/Sep/20 |
|
| Status: | Won't Fix |
| Project: | Instrument control development |
| Component/s: | None |
| Affects Version/s: | None |
| Fix Version/s: | None |
| Type: | Bug | Priority: | Normal |
| Reporter: | arnaud.lefur | Assignee: | arnaud.lefur |
| Resolution: | Won't Fix | Votes: | 0 |
| Labels: | SPS | ||
| Remaining Estimate: | Not Specified | ||
| Time Spent: | Not Specified | ||
| Original Estimate: | Not Specified | ||
| Issue Links: |
|
||||||||||||
| Story Points: | 4 | ||||||||||||
| Sprint: | SM1PD-2020 G, SM1PD-2020 G2 | ||||||||||||
| Description |
|
a flat has failed because red shutter would not open. 2020-06-11 23:52:42.743Z cmds 20 Actor.py:497 new cmd: shutters expose exptime=3.0
2020-06-11 23:52:42.744Z cmds 20 CommandLink.py:121 > 0 0 i biashaFSM=ONLINE,BUSY
2020-06-11 23:52:42.745Z cmds 20 CommandLink.py:121 > 0 0 i metaFSM=ONLINE,BUSY
2020-06-11 23:52:42.745Z biasha 10 bufferedSocket.py:57 sending b'init\r\n'
2020-06-11 23:52:42.754Z biasha 10 bufferedSocket.py:89 received 'ok'
2020-06-11 23:52:42.755Z cmds 20 CommandLink.py:121 > 0 0 i biashaFSM=ONLINE,IDLE 2020-06-11 23:52:42.755Z cmds 20 CommandLink.py:121 > 0 0 i metaFSM=ONLINE,IDLE
2020-06-11 23:52:42.756Z cmds 20 CommandLink.py:121 > 0 0 i biashaFSM=ONLINE,BUSY
2020-06-11 23:52:42.757Z cmds 20 CommandLink.py:121 > 0 0 i metaFSM=ONLINE,BUSY
2020-06-11 23:52:42.758Z biasha 10 bufferedSocket.py:57 sending b'shut_open\r\n'
2020-06-11 23:52:44.780Z biasha 10 bufferedSocket.py:89 received 'shutter switch timeoutnok'
2020-06-11 23:52:44.781Z cmds 20 CommandLink.py:121 > 0 0 i biashaFSM=ONLINE,FAILED
2020-06-11 23:52:44.782Z cmds 20 CommandLink.py:121 > 0 0 i metaFSM=ONLINE,FAILED
2020-06-11 23:52:44.782Z cmds 20 CommandLink.py:121 > 2 1065 w exptime=nan
2020-06-11 23:52:44.790Z cmds 20 CommandLink.py:121 > 2 1065 f text="command failed: RuntimeError('error : shutter switch timeoutn',) in _gotoState() at /software/mhs/products/Linux64/ics_enu
Actor/1.0.4/python/enuActor/Controllers/biasha.py:346"
When I connected shutters was returning this status. pfs@shell-ics:~$ oneCmd.py enu_sm1 shutters status
2020-06-12T11:31:18.847 sent enu_sm1 shutters status
2020-06-12T11:31:18.866 enu_sm1 i biasha="20"
2020-06-12T11:31:18.883 enu_sm1 i shb=1,0,0
2020-06-12T11:31:18.885 enu_sm1 i shr=0,1,0
2020-06-12T11:31:18.887 enu_sm1 w shutters="undef"
2020-06-12T11:31:18.888 enu_sm1 f text="command failed: RuntimeError('statword 100010 does not match current state',) in shutterStatus() at /software/mhs/products/Linux64/ics_enuActor/1.0.4/python/enuAc
tor/Controllers/biasha.py:236"
blue shutter is behaving correctly, but red shutter return an error pfs@shell-ics:~$ oneCmd.py enu_sm1 connect controller=biasha
2020-06-12T11:37:03.783 sent enu_sm1 connect controller=biasha
2020-06-12T11:37:03.925 enu_sm1 i biashafsm=OFF,IDLE
2020-06-12T11:37:03.935 enu_sm1 i metafsm=OFF,IDLE
2020-06-12T11:37:03.938 enu_sm1 i biashafsm=OFF,LOADING
2020-06-12T11:37:03.947 enu_sm1 i metafsm=OFF,LOADING
2020-06-12T11:37:03.950 enu_sm1 i text="biasha configuration correctly Loaded"
2020-06-12T11:37:03.954 enu_sm1 i text="biasha communication is open"
2020-06-12T11:37:03.957 enu_sm1 i biasha="0"
2020-06-12T11:37:03.962 enu_sm1 i text="biasha communication is functional"
2020-06-12T11:37:03.966 enu_sm1 i biashafsm=LOADED,INITIALISING
2020-06-12T11:37:03.968 enu_sm1 i metafsm=LOADED,INITIALISING
2020-06-12T11:37:03.986 enu_sm1 i text="biasha initialisation OK"
2020-06-12T11:37:03.993 enu_sm1 i biashafsm=ONLINE,IDLE
2020-06-12T11:37:03.998 enu_sm1 i biashamode="operation"
2020-06-12T11:37:03.999 enu_sm1 i biasha="0"
2020-06-12T11:37:04.031 enu_sm1 i photores=0,0
2020-06-12T11:37:04.032 enu_sm1 i biaconfig=0,1000,150
2020-06-12T11:37:04.033 enu_sm1 i bia="off"
2020-06-12T11:37:04.049 enu_sm1 i shb=0,1,0
2020-06-12T11:37:04.057 enu_sm1 i shr=0,1,0
2020-06-12T11:37:04.060 enu_sm1 i shutters="close"
2020-06-12T11:37:04.066 enu_sm1 : controllers=temps,rexm,iis,pdu,biasha
pfs@shell-ics:~$ oneCmd.py enu_sm1 shutters blue open
2020-06-12T11:37:18.423 sent enu_sm1 shutters blue open
2020-06-12T11:37:18.448 enu_sm1 f text="Unmatched command: shutters blue open (exception: Invalid cmd: shutters blue open)"
pfs@shell-ics:~$ oneCmd.py enu_sm1 shutters open blue
2020-06-12T11:37:26.086 sent enu_sm1 shutters open blue
2020-06-12T11:37:26.422 enu_sm1 i biasha="30"
2020-06-12T11:37:26.442 enu_sm1 i shb=1,0,0
2020-06-12T11:37:26.445 enu_sm1 i shr=0,1,0
2020-06-12T11:37:26.451 enu_sm1 i shutters="openblue"
2020-06-12T11:37:26.454 enu_sm1 :
pfs@shell-ics:~$ oneCmd.py enu_sm1 shutters close blue
2020-06-12T11:37:42.314 sent enu_sm1 shutters close blue
2020-06-12T11:37:42.657 enu_sm1 i biasha="0"
2020-06-12T11:37:42.677 enu_sm1 i shb=0,1,0
2020-06-12T11:37:42.679 enu_sm1 i shr=0,1,0
2020-06-12T11:37:42.684 enu_sm1 i shutters="close"
2020-06-12T11:37:42.687 enu_sm1 :
pfs@shell-ics:~$ oneCmd.py enu_sm1 shutters open red
2020-06-12T11:37:52.630 sent enu_sm1 shutters open red
2020-06-12T11:37:54.655 enu_sm1 f text="command failed: RuntimeError('error : shutter switch timeoutn',) in _gotoState() at /software/mhs/products/Linux64/ics_enuActor/1.0.4/python/enuActor/Controllers/biasha.py:346"
|
| Comments |
| Comment by rhl [ 12/Jun/20 ] |
|
Was the error returned when the flat failed useful to the innocent observer? |
| Comment by arnaud.lefur [ 12/Jun/20 ] |
text="enu_sm1 shutters failed: RuntimeError('error : shutter switch timeoutn',) in _gotoState() at /software/mhs/products/Linux64/ics_enuActor/1.0.4/python/enuActor/Controllers/biasha.py:346";text="command failed: RuntimeError('exposure failed',) in process() at /software/mhs/products/Linux64/ics_spsActor/1.0.0/python/spsActor/Commands/ExposeCmd.py:79"
shutter switch timeout is an error returned by the arduino control board. Perhaps it could be a bit more specific about what's actually going on ... |
| Comment by hassan [ 10/Sep/20 ] |
|
cloomis and arnaud.lefur report that this is a hardware issue, so no ICS updates required. |