While taking data with HgAr lamp during June/July acquisition I made a mistake and I took the data in the same fashion as with noble gas lamps. Given that the flux of HgAr lamp is a nonlinear function of time, I have seriously overexposed the defocused data, and achieved very low flux in the focused data. This data acquisition has to be repeat, either by using a shutter or by carefully modifying exposure times. During SPS telecon on Wen, August 4 2021, it has been projected that the possibility for this work might exist during end of August, beginning of September.
- blocks
-
PIPE2D-789 Reduce data described in PIPE2D-788
-
- Done
-