Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-80

Ion pump start error: only one started

    XMLWordPrintable

    Details

      Description

      10:25:48.617716 .hub 0 cmds d CmdIn="client.pfs","xcu_r1","ionpump on"
      10:25:48.618379 .hub 0 cmds d CmdQueued=1012496,1484648748.60,"client.pfs",16,"xcu_r1",90,"ionpump on"
      10:25:49.385994 client.pfs 16 xcu_r1 i ionPump1=1,300,2.6e-06,35,1.2e-08
      10:25:50.071941 client.pfs 16 xcu_r1 i ionPump2=0,0,0,35,0
      10:25:50.072684 client.pfs 16 xcu_r1 f text="ion pump controller is in LOCAL mode!"
      10:25:50.073346 .hub 0 cmds d CmdDone=1012496,"f"

      10:26:26.672752 .hub 0 cmds d CmdIn="client.pfs","xcu_r1","ionpump status"
      10:26:26.673473 .hub 0 cmds d CmdQueued=1012539,1484648786.67,"client.pfs",17,"xcu_r1",91,"ionpump status"
      10:26:27.419803 client.pfs 17 xcu_r1 i ionPump1=1,6100,0.013,35,8.5e-05
      10:26:28.146249 client.pfs 17 xcu_r1 i ionPump2=0,0,0,35,0
      10:26:28.146943 client.pfs 17 xcu_r1 :
      10:26:28.147526 .hub 0 cmds d CmdDone=1012539,":"