Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-666

disable holding current on red exchange mechanism

    XMLWordPrintable

    Details

    • Story Points:
      1
    • Sprint:
      SM1-2019 G

      Description

      We need to disable the holding current from rexm to avoid any heat from the motors.

      It's set by two parameters in the motor controller:

      SAP 7, 0, 0 //standby current (0A)

      SAP 214, 0, 20 // power down delay (200ms)

      I just need to add this to the initialisation procedure.

        Attachments

          Activity

            People

            • Assignee:
              arnaud.lefur arnaud.lefur
              Reporter:
              arnaud.lefur arnaud.lefur
            • Votes:
              0 Vote for this issue
              Watchers:
              1 Start watching this issue

              Dates

              • Created:
                Updated:
                Resolved: