Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-631

Understand/handle SAM FIFO timeouts

    XMLWordPrintable

    Details

    • Type: Task
    • Status: Open (View Workflow)
    • Priority: Normal
    • Resolution: Unresolved
    • Component/s: None
    • Labels:

      Description

      Just had this happen, after hundreds of ramps, thousands of reads. Restarting the actor (and thus resetting the ASIC, but not power-cycling anything) fixed the problem. But needs to be figured out. At the very least the timeouts are too long, and a few timeouts in a row should trigger some recovery.

      2019-03-15 21:38:21.075Z sam 40 qusb.py:555 FIFO timeout! need=0x00100000, npix=0x00000006, empty=0x0e, full=0x00

        Attachments

          Activity

            People

            • Assignee:
              Unassigned
              Reporter:
              cloomis cloomis
            • Votes:
              0 Vote for this issue
              Watchers:
              1 Start watching this issue

              Dates

              • Created:
                Updated: