Restore H4RG readouts with new cabling.

    XMLWordPrintable

    Details

    • Type: Task
    • Status: Done (View Workflow)
    • Priority: Blocker
    • Resolution: Done
    • Component/s: None
    • Labels:

      Description

      There are two built up H4 signal chains at JHU: one in the n1 cryostat in the NIR cleanroom, and one in n9 in the Optics lab. Both have new flight cabling: the SAM in is a vacuum feedthrough box, the braided cable harness to the ASIC at the FPA, and the proper ASIC-FPA ribbon. All parts have been tested separately and together.

      But the n9 reads currently look like there is no detector. More specifically, they look like the ASIC is always configured with grounded inputs (bypassing the inputs from the detector).

      This is a warning ticket, without specific work or times.