Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-2349

Update n1 FPA position in xcu.yaml

    XMLWordPrintable

    Details

    • Sprint:
      EngRun18August

      Description

      It was found that n1 was defocused by 53um at Subaru. Update FPA position accordingly

          tilts: [-10, 35, -15]
          focus: 115.0

      focus should be 115.0 - 53 = 62.0