Update threshold of alarm for N4 detector temperature

    Details

    • Sprint:
      preRun18August1

      Description

      ics_alertsActor uses pfs_instdata 1.7.48 for which the n2/n3 h4 detector temperature was adjusted to 101K (note; the original set temperature is 100K).  It was in order to control the detector temperature more easily when SCR room temperature is higher. After it, when n4 was installed in June, n4 detector was set to 101K although alerts was not adjusted. We may have the same issue for n1, which was recently installed.

      So, I think we should either (1) make a new tag-version with all n detector was set to 101K, or (2) change detector temperature to100K. (or do both)

        Attachments

          Activity

          Hide
          cloomis cloomis added a comment -

          I agree with aiming for 101K on all. n2 is the difficult one, where the H4 heater is as low as we want to run it and the ASIC heater is being run quite high in order to try to keep the thing alive:

          cloomis@shell-ics:~$ for cam in n2 n3 n4; do oneCmd.py xcu_$cam heaters status | grep 'heater.=h4'; done
          2024-08-14T19:12:59.196 xcu_n2 i heater2=h4,TEMP,101.00,0.00,0.139,100.9995
          2024-08-14T19:12:59.969 xcu_n3 i heater2=h4,TEMP,101.00,0.00,0.164,100.9994
          2024-08-14T19:13:00.625 xcu_n4 i heater2=h4,TEMP,101.00,0.00,0.220,100.9996
          
          cloomis@shell-ics:~$ for cam in n2 n3 n4; do oneCmd.py xcu_$cam heaters status | grep 'heater.=asic'; done
          2024-08-14T19:12:51.580 xcu_n2 i heater1=asic,TEMP,160.00,0.00,0.691,159.9830
          2024-08-14T19:12:52.177 xcu_n3 i heater1=asic,TEMP,155.00,0.00,0.185,154.9971
          2024-08-14T19:12:52.841 xcu_n4 i heater1=asic,TEMP,155.00,0.00,0.586,154.9970
          
          Show
          cloomis cloomis added a comment - I agree with aiming for 101K on all. n2 is the difficult one, where the H4 heater is as low as we want to run it and the ASIC heater is being run quite high in order to try to keep the thing alive: cloomis@shell-ics:~$ for cam in n2 n3 n4; do oneCmd.py xcu_$cam heaters status | grep 'heater.=h4'; done 2024-08-14T19:12:59.196 xcu_n2 i heater2=h4,TEMP,101.00,0.00,0.139,100.9995 2024-08-14T19:12:59.969 xcu_n3 i heater2=h4,TEMP,101.00,0.00,0.164,100.9994 2024-08-14T19:13:00.625 xcu_n4 i heater2=h4,TEMP,101.00,0.00,0.220,100.9996 cloomis@shell-ics:~$ for cam in n2 n3 n4; do oneCmd.py xcu_$cam heaters status | grep 'heater.=asic'; done 2024-08-14T19:12:51.580 xcu_n2 i heater1=asic,TEMP,160.00,0.00,0.691,159.9830 2024-08-14T19:12:52.177 xcu_n3 i heater1=asic,TEMP,155.00,0.00,0.185,154.9971 2024-08-14T19:12:52.841 xcu_n4 i heater1=asic,TEMP,155.00,0.00,0.586,154.9970
          Hide
          yuki.moritani yuki.moritani added a comment -

          Thank you for the comments. So can we go with 101K for all detectors?

          Show
          yuki.moritani yuki.moritani added a comment - Thank you for the comments. So can we go with 101K for all detectors?
          Hide
          cloomis cloomis added a comment -

          Yes, I think so.

          Show
          cloomis cloomis added a comment - Yes, I think so.
          Hide
          cloomis cloomis added a comment -

          Tagged 1.7.59

          Show
          cloomis cloomis added a comment - Tagged 1.7.59