-
Type: Task
-
Status: Open (View Workflow)
-
Priority: Normal
-
Resolution: Unresolved
-
Component/s: ics_ccdActor, ics_hxActor, ics_utils
-
Labels:None
For object and some calibration exposures we need to use the shutter times. Neither the ccd nor the hx actors do that correctly. And they do it differently. Fix.
cloomis@shell-ics:2024-06-03$ fitsheader -f -e0 -k W_CAM -k HST-STR {sps,ramps}/PFS?111737* filename W_CAM HST-STR ----------------------- ----- ------------ sps/PFSA11173711.fits b1 19:27:22.741 sps/PFSA11173712.fits r1 19:27:22.662 sps/PFSA11173721.fits b2 19:27:46.372 sps/PFSA11173722.fits r2 19:27:46.341 sps/PFSA11173731.fits b3 19:27:40.136 sps/PFSA11173732.fits r3 19:27:40.155 sps/PFSA11173741.fits b4 19:27:22.711 sps/PFSA11173742.fits r4 19:27:22.705 ramps/PFSB11173723.fits n2 19:30:54.448 ramps/PFSB11173733.fits n3 19:30:59.277
(rubin3_ics) alefur@shell-ics:/data/logs/actors$ cat enu_sm*/2024-06-02*|grep shutterTimings=111737 2024-06-03 05:30:48.813Z cmds 20 CommandLink.py:122 > 2 3507 i shutterTimings=111737,2024-06-02T19:27:48.386778,2024-06-02T19:27:48.782632,2024-06-02T19:30:48.387398,2024-06-02T19:30:48.778487 2024-06-03 05:30:48.830Z cmds 20 CommandLink.py:122 > 2 3886 i shutterTimings=111737,2024-06-02T19:27:48.386466,2024-06-02T19:27:48.782507,2024-06-02T19:30:48.387160,2024-06-02T19:30:48.782388 2024-06-03 05:30:48.807Z cmds 20 CommandLink.py:122 > 2 3966 i shutterTimings=111737,2024-06-02T19:27:48.386243,2024-06-02T19:27:48.782049,2024-06-02T19:30:48.387169,2024-06-02T19:30:48.773816 2024-06-03 05:30:48.800Z cmds 20 CommandLink.py:122 > 2 3709 i shutterTimings=111737,2024-06-02T19:27:48.376621,2024-06-02T19:27:48.772964,2024-06-02T19:30:48.377248,2024-06-02T19:30:48.767711