Looks like the first two temp channels on b2 are swapped. There is also some permutation of the first three channels on b4, and/or maybe a dud/poorly bonded sensor:
2023-12-16T00:03:44.842 xcu_b1 : temps=150.6125,275.4944,142.5447,134.4267,277.4743,400.0000,400.0000,400.0000,400.0000,400.0000,162.2954,163.2170 2023-12-16T00:03:45.717 xcu_b2 : temps=276.9311,145.2434,152.1768,135.1048,278.7843,400.0000,400.0000,400.0000,400.0000,400.0000,162.8774,163.1753 2023-12-16T00:03:46.617 xcu_b3 : temps=154.7262,276.8176,152.7764,139.5380,278.6644,400.0000,400.0000,400.0000,400.0000,400.0000,158.2756,158.6268 2023-12-16T00:03:47.668 xcu_b4 : temps=148.3276,208.3578,275.9466,130.7394,277.4626,400.0000,400.0000,400.0000,400.0000,400.0000,162.2418,162.3472
Umm, the first three are the detector box, the Mangin, and the spider box. So I'd expect middle, high, low. Should look at rates to figure out b4.
Swap back in the xcu actor I think. That knows which camera it belongs to, while the SMB program on the pi does not.
But need to figure out the order on b4 first.
- is duplicated by
-
INSTRM-2309 b2 mangin and detectorbox temperature looks flipped
- Open