Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-2028

Avoid pfilamps status collisions by reading new status log

    XMLWordPrintable

    Details

      Description

      The pfilamps allstat command actually invokes that command on the pi. The command actually queries all the hardware. But while running the lamps, the main "process" is also doing that, giving collisions which result in bad data and mangled outputs.

      When adding the new HgCd control I made the main process write to a status log, with a single line per sample. If the status command just reads from the end of that there will be no collisions.

        Attachments

          Activity

            People

            • Assignee:
              cloomis cloomis
              Reporter:
              cloomis cloomis
            • Votes:
              0 Vote for this issue
              Watchers:
              1 Start watching this issue

              Dates

              • Created:
                Updated:
                Resolved: