Type: Task
-
Status: Open (View Workflow)
-
Priority: Normal
-
Resolution: Unresolved
-
Component/s: hxhal, ics_hxActor
-
Labels:None
Both the Teledyne and Loose ASIC firmwares start the ASIC in reset-when-idle mode. This is before there has been any chance to configure the voltages. In many case this will flood the detector with flux which takes a while to burn off, if only because of their persistence.
A small change to the image would start it in do-nothing-while-idle mode, at which point we can set the bias voltages then start the reset-while-idle loop.
We already have a ticket to stop the reset-while-idle loop when preparing to power the ASIC down. We should do both.