Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-2012

requested ramp length shortened in middle of startExposures sequence

    XMLWordPrintable

    Details

    • Type: Bug
    • Status: Won't Fix (View Workflow)
    • Priority: Normal
    • Resolution: Won't Fix
    • Component/s: None
    • Labels:
      None

      Description

      In the middle of a startExposures sequence, the ramp commands sent to hx_n1 lowered the number of reads requested:

      2023-07-06 07:17:36.950Z new cmd: ramp finish exptime=1200.081 obstime=2023-07-05T20:57:33.206832 stopRamp
      2023-07-06 07:18:14.245Z new cmd: ramp nread=113 visit=96389 exptype=object expectedExptime=1200.0
      2023-07-06 07:38:46.012Z new cmd: ramp finish exptime=1200.091 obstime=2023-07-05T21:18:42.394797 stopRamp
      2023-07-06 07:39:23.960Z new cmd: ramp nread=31 visit=96390 exptype=object expectedExptime=1200.0
      2023-07-06 08:00:25.191Z new cmd: ramp nread=31 visit=96391 exptype=object expectedExptime=1200.0
      2023-07-06 08:21:21.666Z new cmd: ramp nread=31 visit=96392 exptype=object expectedExptime=1200.0
      

        Attachments

          Activity