Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-1940

Fix cooling-when-idle loop on pfilamps pi.

    XMLWordPrintable

    Details

      Description

      After a lamp illumination has finished, the cooling fan is turned on for 120s. If a new lamp command comes in before that time ends, the fans can be turned off unexpectedly. Fix the logic to safely cool.

        Attachments

          Activity

            People

            • Assignee:
              cloomis cloomis
              Reporter:
              cloomis cloomis
            • Votes:
              0 Vote for this issue
              Watchers:
              2 Start watching this issue

              Dates

              • Created:
                Updated:
                Resolved: