During last engineering run, some arcs has been taken using gen2:
iic scienceArc hgcd=0.0 argon=0.0 neon=10 krypton=0.0 duplicate=1 cam=b1,r1
Some requested lamp time has been set to 0, I suppose because of how gen2 commands are written, that's fine with me, but we have to be careful about book-keeping.
conda-ics) alefur@shell-ics:~$ fitsheader /data/raw/2022-09-24/sps/PFSA07999012.fits |grep 'W_AIT' W_AITNEO= T / Neon lamp was on W_AITARG= T / Argon lamp was on W_AITKRY= T / Krypton lamp was on W_AITXEN= F / Xenon lamp was on W_AITHGC= T / HgCd lamp was on W_AITQTH= F / Halogen lamp was on (conda-ics) alefur@shell-ics:~$ fitsheader /data/raw/2022-09-24/sps/PFSA07999012.fits |grep 'W_CL' W_CLNEOT= 10 / Neon lamp request time W_CLARGT= 0 / Argon lamp request time W_CLKRYT= 0 / Krypton lamp request time W_CLXENT= 0 / Xenon lamp request time W_CLHGCT= 0 / HgCd lamp request time W_CLQTHT= 0 / Halogen lamp request time
As you can see, for that visit, the flag has been set to True with time=0.
I'm not sure how drp interpret that, but still I think that flag should be set to False
- relates to
-
INSTRM-1727 fixing pfilamps header keys for september22 engRun
- Done