Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-1673

gatevalve open underVacuum on b3 started to open then closed by itself.

    XMLWordPrintable

    Details

      Description

      Within gv_open was received with the turbo at speed and the pressures ok, gv_request was asserted, gv_closed dropped (showing that the GV actually moved), then immediately gv_request was dropped and gv_closed reseen.

      All within 0.4s.

       

      ```2022-07-20T15:58:35.633 xcu_b3 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;gvmask="0xc7";interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed"
      '"22-07-20T15:58:35.634 xcu_b3 d text="sending '?V852
      2022-07-20T15:58:35.697 xcu_b3 d text="recv '=V852 1500;000502BC'"
      2022-07-20T15:58:35.697 xcu_b3 i turbospeed="90000"
      2022-07-20T15:58:35.698 xcu_b3 i turbostatus=0x000502bc,"Above normal speed, Vent valve energized, Start command active, Serial enable active, Above 50% full speed, Serial control mode, bit 17, bit 19"
      2022-07-20T15:58:35.698 xcu_b3 d text="sending '~gStat,all\n'"
      2022-07-20T15:58:35.704 xcu_b3 d text="recv '~gStat,all'"
      2022-07-20T15:58:35.974 xcu_b3 d text="recv '01110100'"
      2022-07-20T15:58:35.974 xcu_b3 d text="sending '~gP,all\n'"
      2022-07-20T15:58:35.979 xcu_b3 d text="recv '~gP,all'"
      2022-07-20T15:58:36.320 xcu_b3 d text="recv '-1.93, -0.38'"
      2022-07-20T15:58:36.320 xcu_b3 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;gvmask="0xc7";interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed"
      2022-07-20T15:58:36.321 xcu_b3 d text="sending '~gStat,all\n'"
      2022-07-20T15:58:36.327 xcu_b3 d text="recv '~gStat,all'"
      2022-07-20T15:58:36.660 xcu_b3 d text="recv '01110100'"
      2022-07-20T15:58:36.661 xcu_b3 d text="sending '~gP,all\n'"
      2022-07-20T15:58:36.665 xcu_b3 d text="recv '~gP,all'"
      2022-07-20T15:58:37.007 xcu_b3 d text="recv '-1.91, -0.41'"
      2022-07-20T15:58:37.008 xcu_b3 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;gvmask="0xc7";interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed"
      2022-07-20T15:58:37.008 xcu_b3 d text="sending '~gStat,all\n'"
      2022-07-20T15:58:37.014 xcu_b3 d text="recv '~gStat,all'"
      2022-07-20T15:58:37.348 xcu_b3 d text="recv '11110001'"
      2022-07-20T15:58:37.349 xcu_b3 i gatevalve=0xf1,unknown,open;gvmask="0xce";interlock=0b11110001,"open_cmd, turbo_at_speed, pressure_equal, vacuum_ok, gv_request"
      2022-07-20T15:58:37.549 xcu_b3 d text="sending '~gStat,all\n'"
      2022-07-20T15:58:37.553 xcu_b3 d text="recv '~gStat,all'"
      2022-07-20T15:58:37.691 xcu_b3 d text="recv '11110000'"
      2022-07-20T15:58:37.691 xcu_b3 i gatevalve=0xf0,unknown,blocked;gvmask="0xcf";interlock=0b11110000,"open_cmd, turbo_at_speed, pressure_equal, vacuum_ok"
      2022-07-20T15:58:37.891 xcu_b3 d text="sending '~gStat,all\n'"
      2022-07-20T15:58:37.896 xcu_b3 d text="recv '~gStat,all'"
      2022-07-20T15:58:38.033 xcu_b3 d text="recv '11110100'"
      2022-07-20T15:58:38.034 xcu_b3 i gatevalve=0xf4,closed,blocked;gvmask="0xcf";interlock=0b11110100,"open_cmd, turbo_at_speed, pressure_equal, vacuum_ok, gv_closed"```

        Attachments

          Activity

            People

            • Assignee:
              shope shope
              Reporter:
              cloomis cloomis
            • Votes:
              0 Vote for this issue
              Watchers:
              1 Start watching this issue

              Dates

              • Created:
                Updated:
                Resolved: