Type: Task
-
Status: Won't Fix (View Workflow)
-
Priority: Normal
-
Resolution: Won't Fix
-
Component/s: None
-
Labels:
-
Story Points:4
-
Sprint:2DDRP-2021 A 6, SM1PD-2021 A 15, SM1PD-2021 A16
(Carrying over from INSTRM-7):
According to a recent report from Jim Gunn https://sumire.pbworks.com/w/file/fetch/145332735/Detectors210616.pdf : a provisional solution to the tearing has been found, that also mitigates the effect on biases. But quoting the summary in that report:
[...]A recent short report from the
Rubin/LSST group, however, showed that the effect varies enough from sensor
to sensor in their set that we have to figure out a way to find the effect in R1, 2, and
3 to be sure it works for all our sensors. This is not trivial in a fiber spectrograph, but
I believe that this can be done by taking a single flat exposure during which we move the slit enough to create a special dithered flat, but this must be tested.
This ticket is to track those tests.
- relates to
-
INSTRM-1481 CCD R4 ensure tearing is no longer problematic
- Done
-
INSTRM-7 CCD flats show regions of vertical structure from top of detector
- Done