Uploaded image for project: 'Instrument control development'
  1. Instrument control development
  2. INSTRM-1210

Assigning NaN or better number to broken fiber locations

    XMLWordPrintable

    Details

      Description

      The broken fiber center locations are assign to be 0+0j in ics_cobraChamer right now. This assignment will create a miss identification to the fiber points. The current fiber identification methods uses nearest-neighborhood. Assigning 0+0j will cause identification of any point close to origin, which is spot of SC14/PID57.

      Also, there is a request from Javier for his simulation work.

        Attachments

          Activity

            People

            • Assignee:
              chyan chyan
              Reporter:
              chyan chyan
              Reviewers:
              hassan
            • Votes:
              0 Vote for this issue
              Watchers:
              3 Start watching this issue

              Dates

              • Created:
                Updated:
                Resolved: