Add an obvious indicator of the optical bench thermal stability

    XMLWordPrintable

    Details

    • Type: Task
    • Status: Open (View Workflow)
    • Priority: Normal
    • Resolution: Unresolved
    • Component/s: None
    • Labels:
    • Story Points:
      3
    • Sprint:
      SM1PD-2020 N, SM1PD-2021 A 12

      Description

      While red shutter was pulled off from SM1, the SCR temperature got an huge gradient which was big enough to affect the thermal stability of the optical bench(see attached plot).
      This has a direct influence on the data itself, so it need to be clearly reported to the user, observer ...