Type:
Bug
-
Status: Won't Fix (View Workflow)
-
Priority:
Blocker
-
Resolution: Won't Fix
-
Component/s: None
-
Labels:
-
Sprint:SM1PD-2020 G
we were unable to open B2 gatevalve :
2020-06-24 12:13:03 cmdIn=xcu_b2 gatevalve open underVacuum 2020-06-24 12:13:03 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 2020-06-24 12:13:04 xcu_b2 i turbospeed="90000" 2020-06-24 12:13:04 xcu_b2 i turbostatus=0x000502bc,"Above normal speed, Vent valve energized, Start command active, Serial enable active, Above 50% full speed, Serial control mode, bit 17, bit 19" 2020-06-24 12:13:04 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 2020-06-24 12:13:05 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 2020-06-24 12:13:05 xcu_b2 i gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 2020-06-24 12:13:09 xcu_b2 w unparsedtext="failed to get desired gate valve state. Timed out with: 0b1110100)" 2020-06-24 12:13:09 xcu_b2 i interlockpressures=0.00,0.00;gatevalve=0x74,closed,closed;interlock=0b1110100,"turbo_at_speed, pressure_equal, vacuum_ok, gv_closed" 2020-06-24 12:13:09 xcu_b2 :
- relates to
-
INSTRM-1028 (re-)Add BEE-side ADIO mask to gatevalve keyword
-
- Done
-